626 Chapter 3
Performance Verification Test Records
Agilent E4411B Performance Verification Test Record
50 dB –0.60 dB (10)________ 0.60 dB ±0.095 dB
55 dB –0.65 dB (11)________ 0.65 dB ±0.099 dB
60 dB –0.70 dB (12)________ 0.70 dB ±0.100 dB
14. Reference Level
Accuracy
Log
Reference Level
50 Ω (dBm) 75 Ω (dBmV)
–15 33.75 –0.3 dB (1)________ 0.3 dB ±0.091 dB
–5 43.75 –0.3 dB (2)________ 0.3 dB ±0.100 dB
–35 13.75 –0.3 dB (3)________ 0.3 dB ±0.092 dB
–45 3.75 –0.3 dB (4)________ 0.3 dB ±0.094 dB
–55 –6.25 –0.5 dB (5)________ 0.5 dB ±0.095 dB
–65 –16.25 –0.5 dB (6)________ 0.5 dB ±0.097 dB
–75 –26.25 –0.7 dB (7)________ 0.7 dB ±0.099 dB
Linear
Reference Level
50Ω (dBm) 75Ω (dBmV)
–15 33.75 –0.3 dB (8)________ 0.3 dB ±0.091 dB
–5 43.75 –0.3 dB (9)________ 0.3 dB ±0.100 dB
–35 13.75 –0.3 dB (10)________ 0.3 dB ±0.092 dB
–45 3.75 –0.3 dB (11)________ 0.3 dB ±0.094 dB
–55 –6.25 –0.5 dB (12)________ 0.5 dB ±0.095 dB
–65 –16.25 –0.5 dB (13)________ 0.5 dB ±0.097 dB
–75 –26.25 –0.7 dB (14)________ 0.7 dB ±0.099 dB
Table 3-16 Agilent E4411B Performance Verification Test Record
Agilent Technologies
Model E4411B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum Results
Measured Maximum Measurement
Uncertainty