Chapter 3 445
Performance Verification Test Records
Agilent E4401B Performance Verification Test Record
53 MHz (1)________ 1.0 dB ±0.123 dB
50.004 MHz
(Option 1DR only)
(2)________ 1.0 dB ±0.123 dB
1403 MHz (3)________ 1.0 dB ±0.123 dB
35. Displayed Average
Noise Level
Note: Enter data in the appropriate section below depending upon the
input impedance and serial number of the analyzer.
50 , 1 kHz RBW, Preamp
Off
400 kHz (1)________ –115 dBm ±1.23 dB
1 MHz to 10 MHz (2)________ –115 dBm ±1.23 dB
10 MHz to 500 MHz (3)________ –119 dBm ±1.23 dB
500 MHz to 1 GHz (4)________ –117 dBm ±1.23 dB
1 GHz to 1.5 GHz (5)________ –114 dBm ±1.23 dB
50 , 1 kHz RBW,
Preamp On
400 kHz (6)________ –131 dBm ±1.23 dB
1 MHz to 10 MHz (7)________ –131 dBm ±1.23 dB
10 MHz to 500 MHz (8)________ –135 dBm ±1.23 dB
500 MHz to 1 GHz (9)________ –133 dBm ±1.23 dB
1 GHz to 1.5 GHz (10)________ –131 dBm ±1.23 dB
50 , 10 Hz RBW,
Preamp Off
400 kHz (11)________ –134 dBm ±1.23 dB
1 MHz to 10 MHz (12)________ –134 dBm ±1.23 dB
10 MHz to 500 MHz (13)________ –138 dBm ±1.23 dB
500 MHz to 1 GHz (14)________ –136 dBm ±1.23 dB
1 GHz to 1.5 GHz (15)________ –133 dBm ±1.23 dB
Table 3-2 Agilent E4401B Performance Verification Test Record
Agilent Technologies
Model E4401B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum Results
Measured Maximum Measurement
Uncertainty