472 Chapter  3 
Performance Verification Test Records
Agilent E4402B Performance Verification Test Record
1kHz RBW, Preamp On,
20 to 30° C
          10 MHz to 1 GHz (13)________ –133 dBm ±1.23 dB
          1 GHz to 2 GHz (14)________ –134 dBm ±1.23 dB
          2 GHz to 3 GHz (15)________ –132 dBm ±1.23 dB
10 Hz RBW, Preamp On, 
20 to 30° C
          10 MHz to 1 GHz (16)________ –152 dBm ±1.23 dB
          1 GHz to 2 GHz (17)________ –152 dBm ±1.23 dB
          2 GHz to 3 GHz (18)________ –151 dBm ±1.23 dB
39. Residual Responses
150 kHz to 3.0 GHz (1)________ –90 dBm ±0.87 dB
40. Fast Time Domain 
Amplitude Accuracy 
(Option AYX only)
  Amplitude Error –0.3% (1)________ 0.3% ±0.029%
42. Tracking Generator 
Absolute Amplitude 
and Vernier Accuracy 
(Option 1DN only)
Absolute Amplitude 
Accuracy
–0.75 dB (1)________ 0.75 dB ±0.096 dB
Absolute Vernier 
Accuracy
–2 dB –0.5 dB (2)________ 0.5 dB ±0.008 dB
–3 dB –0.5 dB (3)________ 0.5 dB ±0.008 dB
–5 dB –0.5 dB (4)________ 0.5 dB ±0.008 dB
–6 dB –0.5 dB (5)________ 0.5 dB ±0.008 dB
–7 dB –0.5 dB (6)________ 0.5 dB ±0.008 dB
Table 3-4  Agilent E4402B Performance Verification Test Record
Agilent Technologies
Model E4402B Report No. ___________ 
Serial No. ___________ Date ___________
Test Description Minimum Results 
Measured Maximum Measurement 
Uncertainty