
126 Chapter 2 
Performance Verification Tests
16. Resolution Bandwidth Switching Uncertainty
Input/Output (or Input), Amptd Ref (On) (Agilent E4401B, E4411B)
Input/Output (or Input), Amptd Ref Out (On) (Agilent E4402B, 
E4403B, E4404B, E4405B, E4407, E4408B)
FREQUENCY, Center Freq, 50 MHz
SPAN, 5kHz
AMPLITUDE,More, Y Axis Units (or Amptd Units), dBm 
AMPLITUDE, –25 dBm (Agilent E4401B, E4411B)
AMPLITUDE, –19 dBm (Agilent E4402B, E4403B, E4404B, E4405B, 
E4407B, E4408B)
AMPLITUDE, Scale/Div, 1dB
BW/Avg, 1kHz 
BW/Avg, Video BW, 300 Hz 
3. Press AMPLITUDE and use the knob to adjust the reference level until 
the signal appears five divisions below the reference level, then press 
the following keys: 
Peak Search (or Search)
Marker→, Mkr →CF
Marker, Delta
4. Set the analyzer span and resolution bandwidth according to 
Table 2-28. 
5. Press Peak Search (or Search), Marker→, Mkr →CF, Peak Search (or 
Search) then record the marker delta amplitude reading in the 
performance verification test record as indicated in Table 2-28. 
6. Repeat step 4 and step 5 for each of the remaining resolution 
bandwidth and span settings listed in Table 2-28. 
Table 2-28  Resolution Bandwidth Switching Uncertainty 
Analyzer Settings
Marker Delta 
Amplitude 
Reading
RES BW SPAN  Test Record Entry 
1 kHz 5 kHz 0 (Ref)
3 kHz 10 kHz 1)
9 kHza50 kHz 2)
10 kHz 50 kHz 3)
30 kHz 100 kHz 4)
100 kHz 500 kHz 5)
120 kHza500 kHz 6)
300 kHz 1 MHz 7)