110 Chapter 2
Performance Verification Tests
14. Reference Level Accuracy: Agilent E4401B and E4411B
amplitude and enter the result as the Test Record entry in the
performance test record.
The following is an example for –35 dBm reference level:
Analyzer marker amplitude 10.17() dB=
Actual Attenuator Step 30 dB()10.07() dB=
Test Record Entry 10.17() dB 10.07() dB 0.10() dB==