Chapter 3 539
Performance Verification Test Records
Agilent E4405B Performance Verification Test Record
120 kHz –0.3 dB (6)________ 0.3 dB ±0.048 dB
300 kHz –0.3 dB (7)________ 0.3 dB ±0.048 dB
1 MHz –0.3 dB (8)________ 0.3 dB ±0.048 dB
3 MHz –0.3 dB (9)________ 0.3 dB ±0.048 dB
5 MHz –0.6 dB (10)________ 0.6 dB ±0.076 dB
300 Hz (Option 1DR only) –0.3 dB (11)________ 0.3 dB ±0.048 dB
200 Hz (Option 1DR only) –0.3 dB (12)________ 0.3 dB ±0.048 dB
100 Hz (Option 1DR only) –0.3 dB (13)________ 0.3 dB ±0.048 dB
30 Hz (Option 1DR only) –0.3 dB (14)________ 0.3 dB ±0.048 dB
10 Hz (Option 1DR only) –0.3 dB (15)________ 0.3 dB ±0.048 dB
3 Hz (Option 1DR and
1D5 only) Firmware
Revision A.08.00 or later
–0.3 dB (16)________ 0.3 dB ±0.048 dB
1Hz (Option 1DR and
1D5 only) Firmware
Revision A.08.00 or later
–0.3 dB (17)________ 0.3 dB ±0.048 dB
18. Absolute Amplitude
Accuracy (Reference
Settings)
Log, Preamp Off –0.34 dB (1)________ 0.34 dB ±0.09 dB
Lin, Preamp Off –0.34 dB (2)________ 0.34 dB ±0.09 dB
Log, Preamp On
(Option 1DS only)
–0.37 dB (3)________ 0.37 dB ±0.09 dB
Lin, Preamp On
(Option 1DS only)
–0.37 dB (4)________ 0.37 dB ±0.09 dB
20. Overall Absolute
Amplitude Accuracy
0 dBm Reference Level
0 dBm input –0.54 dB (1)________ 0.54 dB ±0.234 dB
Table 3-10 Agilent E4405B Performance Verification Test Record
Agilent Technologies
Model E4405B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum Results
Measured Maximum Measurement
Uncertainty