Chapter  3 529
Performance Verification Test Records
Agilent E4405B Performance Verification Test Record
       20 kHz (2)________ –100 dBc/Hz ±2.44 dB
       30 kHz (3)________ –106 dBc/Hz ±2.44 dB
 (Serial Prefix ≥ US4510, 
SG4510 or MY4510, 
Option 1DR)
       10 kHz  (1)________ –98 dBc/Hz ±2.44 dB
       20 kHz (2)________ –104 dBc/Hz ±2.44 dB
       30 kHz (3)________ –110 dBc/Hz ±2.44 dB
 (Serial Prefix ≥ US4510, 
SG4510 or MY4510, 
Non-Option 1DR)
       10 kHz  (1)________ –90 dBc/Hz ±2.44 dB
       20 kHz (2)________ –104 dBc/Hz ±2.44 dB
       30 kHz (3)________ –110 dBc/Hz ±2.44 dB
8. Noise Sidebands - Wide 
Offsets
Note: Enter data in the appropriate section below depending upon 
whether or not Option 120 (ACPR Dynamic Range Extension) is 
installed. 
 Non-Option 120 
 Offset from 1 GHz signal 
       100 kHz  (1)________ –118 dBc/Hz ±2.44 dB
       1 MHz (2)________ –125 dBc/Hz ±2.44 dB
       5 MHz (3)________ –127 dBc/Hz ±2.44 dB
       10 MHz (4)________ –131 dBc/Hz ±2.44 dB
 Option 120 
 Offset from 1 GHz signal 
       100 kHz  (1)________ –118 dBc/Hz ±2.44 dB
       1 MHz (2)________ –133 dBc/Hz ±2.44 dB
       5 MHz (3)________ –135 dBc/Hz ±2.44 dB
Table 3-10  Agilent E4405B Performance Verification Test Record
Agilent Technologies
Model E4405B Report No. ___________ 
Serial No. ___________ Date ___________
Test Description Minimum Results 
Measured Maximum Measurement 
Uncertainty