Chapter 3 551
Performance Verification Test Records
Agilent E4405B Performance Verification Test Record
–2 dB –0.5 dB (2)________ 0.5 dB ±0.008 dB
–3 dB –0.5 dB (3)________ 0.5 dB ±0.008 dB
–5 dB –0.5 dB (4)________ 0.5 dB ±0.008 dB
–6 dB –0.5 dB (5)________ 0.5 dB ±0.008 dB
–7 dB –0.5 dB (6)________ 0.5 dB ±0.008 dB
–8 dB –0.5 dB (7)________ 0.5 dB ±0.008 dB
–9 dB –0.5 dB (8)________ 0.5 dB ±0.008 dB
–10 dB –0.5 dB (9)________ 0.5 dB ±0.008 dB
Incremental Vernier
Accuracy
–2 dB –0.2 dB (10)________ 0.2 dB ±0.008 dB
–3 dB –0.2 dB (11)________ 0.2 dB ±0.008 dB
–5 dB –0.2 dB (12)________ 0.2 dB ±0.008 dB
–6 dB –0.2 dB (13)________ 0.2 dB ±0.008 dB
–7 dB –0.2 dB (14)________ 0.2 dB ±0.008 dB
–8 dB –0.2 dB (15)________ 0.2 dB ±0.008 dB
–9 dB –0.2 dB (16)________ 0.2 dB ±0.008 dB
–10 dB –0.2 dB (17)________ 0.2 dB ±0.008 dB
44. Tracking Generator
Level Flatness (Option
1DN only)
Positive Level Flatness,
<1 MHz
(1)________ 3.0 dB ±0.284 dB
Negative Level Flatness,
<1 MHz
–3.0 dB (2)________ ±0.284 dB
Positive Level Flatness,
1 MHz to 10 MHz
(3)________ 3.0 dB ±0.142 dB
Table 3-10 Agilent E4405B Performance Verification Test Record
Agilent Technologies
Model E4405B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum Results
Measured Maximum Measurement
Uncertainty