Chapter 2 179
Performance Verification Tests
23. Frequency Response, Agilent E4402B and E4403B
For example, if Flatness Error is –0.30 dB and 50 MHz Ref Amptd is 
0.15 dB, Flatness Relative to 50 MHz would be –0.45 dB.
31.If the analyzer has Option UKB, continue with this procedure. 
Otherwise, perform the procedure in “Measuring Frequency 
Response Less Than or Equal to 100 kHz.”
32.Replace the function generator with the synthesized sweeper.
33.Set the source frequency to 10 MHz and the source power level to the 
value corresponding to the source power setting in Table 2-44 for the 
current source frequency (10 MHz).
34.For an Agilent E4402B only, press Input/Output (or Input), Coupling 
(AC) on the analyzer.
35.For an Agilent E4402B only, repeat step 5 through step 30 for ac 
coupling.
36. Perform the procedure in “Measuring Frequency Response Less 
Than or Equal to 100 kHz.”
Table 2-45  Frequency Response Worksheet, 100 kHz to 3.0 GHz
Freq.
(MHz)
Splitter 
Tracking 
Error
Current
Channel B
Reading
Marker (Mkr1) 
Amptd Flatness Error
Flatness
Relative to
50 MHz
dc ac dc ac dc ac dc ac
0.1
0.5
1 
5 
10a 
10b 
20 
50  0 dB 
(ref)
0 dB 
(ref)
75 
175 
275 
375 
475