300 Chapter 2
Performance Verification Tests
35. Displayed Average Noise Level: Agilent E4401B and E4411B
Figure 2-53 Displayed Average Noise Level Test Setup
2. Press System, More, Show System. If there is an entry shown which
reads 1DR: Narrow Resolution BW and the analyzer is an E4401B,
then enter 10 Hz below as the Minimum RBW. If there is an entry
shown which reads 1DR: Narrow Resolution BW and the analyzer
is an E4411B, then enter 100 Hz below as the Minimum RBW.
Otherwise, enter 1 kHz as the Minimum RBW (resolution
bandwidth). Keep this value in mind as you continue to perform this
procedure. Also, take note of whether or not Options 1DN or 1DQ
(1.5 GHz Tracking Generator) and Option 1DS (RF Preamplifier) are
installed.
Minimum RBW____________ Hz
Option 1DN or 1DQ: ____________
Option 1DS: ____________
3. On the analyzer, press Preset. Press the Factory Preset softkey, if it is
displayed. Then press the following keys:
Input/Output (or Input), Amptd Ref (On)
FREQUENCY, Center Freq, 50 MHz
SPAN, 2kHz
AMPLITUDE, –25 dBm (50 Input only)
AMPLITUDE, 28.75 dBmV (75 Input only)
AMPLITUDE, Attenuation, 10 dB
BW/Avg, Res BW, 1 kHz
BW/Avg, Video BW, 1 kHz
Det/Demod, Detector, Sample
4. On the analyzer, press Single, Peak Search (or Search) and record the
Ref Amptd reading below.
Ref Amptd ____________ dBm (50 Input only)
Ref Amptd ____________ dBmv (75 Input only)
5. If the analyzer does not have Option 1DN or 1DQ (1.5 GHz Tracking