Chapter 2 251
Performance Verification Tests
29. Other Input-Related Spurious Responses: Agilent E4402B, E4403B, E4404B, E4405B,
E4407B, and E4408B
10.Repeat step a through step h using the data in Table 2-68 for Band 0.
a. Set the synthesized sweeper to the listed CW frequency.
Do not set the synthesized sweeper to frequencies outside the
frequency range of the analyzer.
b. Enter the appropriate power sensor calibration factor into the
power meter.
c. Adjust the synthesized sweeper power level until the power meter
reading is equal to the Mixer Level in Table 2-68, ±0.1 dB.
d. On the analyzer, press Single and wait for the completion of a new
sweep. Then, press Peak Search (or Search).
e. If the Mixer Level in Table 2-68 is –20 dBm, press Marker,
Select Marker (2).
f. If the Mixer Level in Table 2-68 is –10 dBm, press Marker,
Select Marker (1).
g. On the analyzer, press Peak Search (or Search).
h. Record the delta marker ( Mkr) amplitude reading in Table 2-68
and in the appropriate entry locations in the performance
verification test record.
11.On the analyzer, press the following keys:
Marker, More, Marker All Off
Auto Couple
SPAN, 1 MHz
AMPLITUDE, Ref Level, –10 dBm
AMPLITUDE, Attenuation, 0 dB
Sweep, Sweep (Cont)
NOTE End of procedure for Agilent E4402B and E4403B.
Band 1
12.On the analyzer, press the following keys:
FREQUENCY, Center Freq, 4 GHz
13.On the synthesized sweeper, press CW, 4 GHz.
14.Enter the power sensor 4 GHz calibration factor into the power
meter.
15.On the analyzer, press the following keys:
Marker, More, Marker All Off
Amplitude, Presel Center
16.Repeat step 5 through step 11 for the synthesized sweeper CW