26 Chapter 1
Calibrating
Before You Start
Oscilloscope Bandwidth: dc to 100 MHz
Vertical Scale Factor of 0.5 V to 5 V/Div
Two channels
Minimum Timebase Setting: <100 ns
Digitizing display with pulse width and
time interval measurement functions
Delta –T measurement accuracy in
200 ns / div: <450 ps
54820A T
Power Meter Compatible with HP/Agilent 8480
series power sensors.
dB relative mode.
Resolution: 0.01 dB
Reference Accuracy: ±1.2%
E4419B P,A,T
RF Power Sensor
(2 required)
Frequency Range: 100 kHz to 3 GHz
Maximum SWR:
1.60 (100 kHz to 300 kHz)
1.20 (300 kHz to 1 MHz)
1.1 (1 MHz to 2.0 GHz)
1.18 (2.0 GHz to 3.0 GHz)
Amplitude range: –25 dBm to 10 dBm
8482A P,A,T
Microwave Power
Sensor
Frequency Range: 50 MHz to 26.5 GHz
Maximum SWR:
1.15 (50 MHz to 100 MHz)
1.10 (100 MHz to 2 GHz)
1.15 (2 GHz to 12.4 GHz)
1.20 (12.4 GHz to 18 GHz)
1.25 (18 GHz to 26.5 GHz)
Amplitude range: –25 dBm to 0 dBm
8485A P,A,T
Power Sensor, Low
Power
Frequency Range: 50 MHz to 3.0 GHz
Amplitude Range: –20 dBm to –70 dBm
Maximum SWR:
1.4 (10 MHz to 30 MHz)
1.15 (30 MHz to 3.0 GHz)
8481D P,A,T
Synthesized Signal
Generator
Frequency Range: 100 kHz to
1500 MHz
Amplitude Range: –35 to 16 dBm
SSB Noise: <–120 dBc/Hz at 20 kHz
offset
8663A P,A
Table 1-9 Recommended Test Equipment
Equipment Critical Specifications for Equipment
Substitution
Recommended
HP/Agilent Model Use a