122 Chapter 2 
Performance Verification Tests
15. Reference Level Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and 
E4408B.
AMPLITUDE, Scale Type (Lin)
AMPLITUDE, More, Y Axis Units (or Amptd Units), dBm 
AMPLITUDE, –20 dBm 
SPAN, 50 kHz
BW/Avg, 3 kHz
Video BW, 30 Hz
Sweep, Sweep Cont
Marker, Off
4. Set the 1 dB step attenuator to place the signal peak 1 to 3 divisions 
below the reference level. 
5. On the analyzer, press the following keys: 
Single
Peak Search (or Search)
Marker, Delta
6. Set the 10 dB step attenuator and analyzer reference level according 
to Column 1 and Column 4 of Table 2-26. At each setting, do the 
following:
a. Press Single on the analyzer.
b. Press Peak Search (or Search).
c. Record the marker delta amplitude reading in Column 5 of 
Table 2-26.
d. Add the Actual Attenuator Step to the Analyzer Marker Delta 
Amplitude and enter the result as the Test Record entry in the 
performance test record.