Main              
  Calibration Guide
Agilent Technologies ESA Spectrum Analyzers               
Notice
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
              Calibration Cycle 
Performance Verification Test Tables               
Table 1-1  Agilent E4401B Performance Verification Tests
            Page
              Table 1-2  Agilent E4402B Performance Verification Tests
              Table 1-3  Agilent E4403B Performance Verification Tests
            Page
              Table 1-4  Agilent E4404B Performance Verification Tests
            Page
              Table 1-5  Agilent E4405B Performance Verification Tests
            Page
              Table 1-6  Agilent E4407B Performance Verification Tests
              Table 1-7  Agilent E4408B Performance Verification Tests
              Table 1-8  Agilent E4411B Performance Verification Tests
              Before You Start 
Recording the test results                 
Performing Self-Alignment 
              Periodic verification of operation 
Test equipment you will need             
Page
              < < < <
            Page
            Page
            Page
              Table 1-11  Recommended Adapters 
              Table 1-12  Recommended Cables 
            Page
              Tests included in this section:
            Page
            Page
              Calibration
            Page
              Table 2-1  Calibration Requirements
            Page
              Table 2-2  Agilent 11713A Settings for Agilent 8494G and Agilent 8496G
              1. 10 MHz Reference Output Accuracy
            Page
            Page
            Page
            Page
              5 Minute Warm-up Error  Counter Reading 1 Counter Reading 3()10=
15 Minute Warm-up Error  Counter Reading 2 Counter Reading 3()10=            
Page
            Page
            Page
            Page
            Page
            Page
            Page
              5. Frequency Span Accuracy: Agilent E4401B  and E4411B
              100 kHz and 100 MHz Frequency Span Readout  Accuracy 
            Page
              Table 2-8  Frequency Span Readout Accuracy 
              6. Frequency Span Accuracy:  Agilent E4402B, E4403B, E4404B, E4405B, 
            Page
            Page
              Table 2-9  Frequency Span Readout Accuracy 
              7. Noise Sidebands
            Page
            Page
            Page
            Page
            Page
            Page
              8. Noise Sidebands - Wide Offsets: Agilent  E4402B, E4404B, E4405B and E4407B 
            Page
            Page
            Page
            Page
              9. System-Related Sidebands
Equipment Required                 
Additional Equipment for 75  Input
Procedure             
Page
            Page
              10. Residual FM
              Procedure 
Part 1: Residual FM             
Page
              Part 2: Residual FM for Option 1DR and 1D5
            Page
              11. Sweep Time Accuracy
            Page
              Sweep Time Accuracy 100 Mkr1 0.8 Sweep Time() Sweep Time
------------------------------------------------------------------------              
12. Display Scale Fidelity
            Page
              Part 1: Log Display Scale Fidelity, Analog Bandwidths
              CLFE Total Actual Attenuation Mkr Reading Total Actual Atten  0 dB from Ref Level()+=
              90 Chapter 2 
Column  8                
Column  9
Test  Record  Entry   CLFE (dB)                
Test  Record  Entry   ILFE  (dB)
              Part 2: Log Display Scale Fidelity, Digital Bandwidths
92 Chapter 2                 
BW/Avg, Average (On) 5, Enter
              CLFE Total Actual Attenuation MkrReading Total Actual Atten
 0 dB from R                
(+=
Chapter 2 93              
Part 3: Linear Display Scale Fidelity, Analog  Bandwidths
System, Alignments, Auto Align, Off FREQUENCY, 50 MHz                
ILFE CLFE current()CLFE previous()=
              Ideal Mkr Reading mV()1000 0.05 10ATmeasATref+()10
Ideal Mkr Reading(mV) 1000 0.05 107.9820.012+()1089.3 mV==              
Part 4: Linear Display Scale Fidelity, Digital  Bandwidths
This section is for analyzers with Option 1DR (narrow resolution  bandwidths) only.                
LFE % of RL()100 Actual Mkr Reading Ideal Mkr Reading 223.6 mV
----------------------------------------------------------------------------------------------------                
Actual Mkr Reading 85.0 mV=  Ideal Mkr Reading 89.3 mV= LFE %of RL()100 85.0 89.3 223.6
              Ideal Mkr Reading mV()1000 0.05 10ATmeasATref+()10
LFE %of RL()100 Actual Mkr Reading Ideal Mkr Reading 223.6 mV                
----------------------------------------------------------------------------------------------------
Chapter 2 97              
10.If the analyzer has a 75W Input press:
Input/Output (or Input) Input Z Corr (75)                
15. On the analyzer press Peak Search (or Search), Marker >, and Marker > CF.
Table 2-16  Linear Scale Fidelity Worksheet, Option 1DR Narrow  Resolution Bandwidths            
Page
              Table 2-17  Zero Span Cumulative Log Fidelity Worksheet, Opt 1DR Narrow  Resolution Bandwidths
              13. Input Attenuation Switching Uncertainty
            Page
              Table 2-18  Actual Attenuation Worksheet
              A
              7.   Preset the synt  hesized signal generato  r ( set the controls as follows:
            Page
            Page
              14. Reference Level Accuracy: Agilent E4401B  and E4411B
              Actual Attenuator Step  X dB()Actual Attenuation 20 dB()Actual Attenuation  X dB()()=
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
              15. Reference Level Accuracy: Agilent E4402B, E4403B, E4404B, E4405B, E4407B, and E4408B.
              Actual Attenuator Step  X dB()Actual Attenuation 20 dB()Actual Attenuation  X dB()()=
            Page
              performance test record. The following is an example for 30 dBm reference level:
Log Scale, Digital Bandwidths, Option 1DR                  
8. Set the 1 dB step attenuator to set the signal peak 1 to 3 divisions  below the reference level.
Table 2-24  Log Mode, Analog Bandwidths Worksheet            
Page
            Page
            Page
              Table 2-27  Linear Mode, Digital Bandwidths Worksheet, Option 1DR
              16. Resolution Bandwidth Switching  Uncertainty
Equipment Required for Agilent E4402B, E4403B,  E4404B, E4405B, E4407B, and E4408B            
Page
            Page
            Page
              17. Absolute Amplitude Accuracy (Reference  Settings): Agilent E4401B and E4411B
            Page
            Page
            Page
            Page
            Page
            Page
              18. Absolute Amplitude Accuracy (Reference  Settings): Agilent E4402B, E4403B, E4404B, 
            Page
            Page
            Page
            Page
            Page
            Page
              19. Overall Absolute Amplitude Accuracy: 
            Page
            Page
              Amplitude Accuracy   Measured Amplitude   Amptd 
            Page
            Page
            Page
              20. Overall Absolute Amplitude Accuracy:  Agilent E4402B, E4403B, E4404B, E4405B, 
            Page
            Page
              Amplitude Accuracy   Measured Amplitude   Amptd 
            Page
            Page
            Page
              21. Resolution Bandwidth Accuracy
            Page
            Page
            Page
              6 dB Bandwidth Upper Marker Frequency Lower Marker Frequency=
              22. Frequency Response: Agilent E4401B and  E4411B
            Page
            Page
              Table 2-40  Source/Splitter Characterization
a. These values do not apply to analyzers with 75  inputs  (Option 1DP).            
Page
            Page
              a. These values do not apply to analyzers with 75  inputs (Option 1DP).
            Page
            Page
            Page
              23. Frequency Response, Agilent E4402B and  E4403B
            Page
            Page
              Table 2-44  Source/Splitter Characterization
            Page
            Page
            Page
            Page
              Table 2-45  Frequency Response Worksheet, 100 kHz to 3.0 GHz
              Table 2-45  Frequency Response Worksheet, 100 kHz to 3.0 GHz
            Page
            Page
            Page
            Page
            Page
              24. Frequency Response, Agilent E4404B, 
            Page
            Page
              Table 2-49  Source/Splitter Characterization
            Page
            Page
            Page
            Page
              Table 2-50  Frequency Response Worksheet, 100 kHz to 3.0 GHz
              Table 2-50  Frequency Response Worksheet, 100 kHz to 3.0 GHz
            Page
            Page
              a. These frequencies apply only to analyzers with  Option UKB.
Table 2-51  Frequency Response Worksheet, 100 kHz            
Page
            Page
            Page
            Page
              Table 2-52  Frequency Response Worksheet, >3 GHz
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
              25. Frequency Response (Preamp On):  Agilent E4401B
            Page
            Page
            Page
            Page
            Page
            Page
            Page
              26. Frequency Response (Preamp On):  Agilent E4402B
            Page
            Page
            Page
            Page
            Page
            Page
              Table 2-61  Frequency Response Worksheet, Preamp On
            Page
            Page
              27. Frequency Response (Preamp On):  Agilent E4404B, E4405B, and E4407B
            Page
            Page
              a. This entry is for data taken with the function  generator as source.
Table 2-63  Source/Splitter Characterization            
Page
            Page
            Page
            Page
              Table 2-64  Frequency Response Worksheet, Preamp On
            Page
            Page
            Page
            Page
              28. Other Input-Related Spurious Responses: 
            Page
            Page
            Page
              Table 2-67  Other Input-Related Spurious Responses Worksheet
a. Image response b. Multiple response            
Page
            Page
            Page
            Page
              30.On the analyzer, press AMPLITUDE, Presel Center. 31.On the analyzer, press the following: 
                 a. Image response b. Multiple response c. Out-of-band response
Table 2-68  Other Input-Related Spurious Responses                 
   
              30. Spurious Responses: Agilent E4401B and  E4411B
            Page
            Page
            Page
            Page
            Page
            Page
            Page
              31. Spurious Responses: Agilent E4402B and  E4403B
            Page
            Page
            Page
            Page
            Page
              0.60 dB 7.05 dBm 6.45 dBm()=
              0.70 dB 7.05 dBm 6.35 dBm()=
              38.4 dBm 20 dBm 59 dB() 0.60 dB()+=
              52.3 dBm 20 dBm 73 dB() 0.70 dB()+=
              32. Spurious Responses: Agilent E4404B, 
            Page
            Page
            Page
            Page
            Page
            Page
            Page
              0.60 dB 7.05 dBm 6.45 dBm()=
              0.70 dB 7.05 dBm 6.35 dBm()=
              1.2 dB 7.35 dBm 6.15 dBm()=
              1.2 dB 7.25 dBm 6.05 dBm()=
            Page
            Page
            Page
            Page
            Page
              33. Gain Compression: Agilent E4401B,  E4402B, E4403B, and E4411B
            Page
              Table 2-77  Source Frequency and Amplitude Settings
a. Agilent E4401B and E4402B with Option 1DR only. b. Agilent E4402B and E4403B only.             
Page
              34. Gain Compression: Agilent E4404B, 
            Page
              Table 2-80 for each frequency indicated. Table 2-79  Source Frequency and Amplitude Settings
            Page
            Page
              35. Displayed Average Noise Level: 
            Page
              Ref Lvl Offset (1 kHz RBW) Ref Amptd Meas Amptd (1 kHz RBW)=
            Page
            Page
              Table 2-81  DANL Measurement Sequence, E4401B 
Table 2-82  DANL Measurement Sequence, E4411B               
Table 2-83  DANL Measurement Sequence, E4401B 75  Inputs
Table 2-82  DANL Measurement Sequence, E4411B             
Page
            Page
            Page
              36. Displayed Average Noise Level: Agilent  E4402B and E4403B
            Page
              Ref Lvl Offset(1 kHz RBW) Ref Amptd Meas Amptd(1 kHz RBW)=
            Page
              Subset D in Table 2-85.
8. After performing all applicable DANL measurement procedures,                 
Table 2-85  DANL Measurement Sequence, E4402B
            Page
            Page
              37. Displayed Average Noise Level: Agilent  E4404B and E4405B
            Page
              Ref Lvl Offset (1 kHz RBW) Ref Amptd Meas Amptd (1 kHz RBW)=
Ref Lvl Offset (10 Hz RBW) Ref Amptd Meas Amptd (10 Hz RBW)=            
Page
              Table 2-87  DANL Measurement Sequence, E4404B
            Page
            Page
              38. Displayed Average Noise Level: Agilent  E4407B and E4408B
            Page
              Ref Lvl Offset(1 kHz RBW)  Ref Amptd Meas Amptd(1 kHz RBW)=
            Page
              Subset D in Table 2-89. Table 2-89  DANL Measurement Sequence, E4407B
            Page
            Page
              39. Residual Responses
            Page
            Page
            Page
              Amplitude Error Mkr1 1.0()100= Amplitude Error Mkr1 100= %
              Equipment Required 
Additional Equipment for Option 1DQ            
Page
              Vernier Accuracy Measured Power Level (dB) Source Vernier Setting (dB)=
Power Sweep Accuracy Positive Power Sweep Accuracy Negative Power Sweep Accuracy=              
Table 2-92  Vernier Accuracy Worksheet
            Page
            Page
            Page
              Vernier Accuracy Measured Power Level (dB) Source Vernier Setting (dB)( 4 dB )+=
              43. Tracking Generator Level Flatness:  Agilent E4401B and E4411B 
dBm 10 log10 E2R()1mW()=              
Additional Equipment for Option 1DQ
Procedure  Tracking Generator Level Flatness, Center  Frequency 100 kHz            
Page
            Page
              Tracking Generator Level Flatness, Center  Frequency 100 kHz
              DVM Offset  DVM Readout= Level Flatness
Corrected Level Flatness  DVM= DVM Offset+            
Page
              dBm 10 log10 E2R()1mW()=
              Tracking Generator Level Flatness, Center  Frequency  100 kHz
            Page
              Tracking Generator Level Flatness, Center  Frequency 100 kHz
              DVM Offset DVM Readout= Level Flatness
Corrected Level Flatness DVM Readout= DVM Offset+            
Page
              45. Tracking Generator Harmonic Spurious  Outputs: Agilent E4401B and E4411B 
Equipment Required                 
Additional Equipment Required for Option 1DQ
Procedure             
Page
            Page
            Page
            Page
            Page
            Page
            Page
              47. Tracking Generator Non-Harmonic  Spurious Outputs: Agilent E4401B and E4411B 
            Page
              Measuring Fundamental Amplitudes
Measuring Non-Harmonic Responses             
Page
              Non-harmonic Amplitude Marker Amplitude Fundamental Amplitude=
              Determining the Highest Non-harmonic Spurious  Response
            Page
            Page
            Page
              Non-harmonic Amplitude Marker Amplitude Fundamental Amplitude=
            Page
            Page
            Page
            Page
            Page
            Page
              Figure 2-70  Gate Delay and Gate Length Test Setup 
3. Set up the function generator as follows:                
4.   Press  
            Page
            Page
              Table 2-106  Gate Delay and Gate Length Accuracy Worksheet
              Additional Equipment for Option 1DP
            Page
            Page
              52. First LO OUTPUT Power Accuracy (Option  AYZ only)
            Page
              Table 2-107  First LO Output Amplitude Accuracy Worksheet
              53. IF INPUT Accuracy (Option AYZ only)
            Page
              New CW Fre
q                
uenc
y                
321.4 MHz Mkr1 Fre
            Page
              54. Comms Frequency Response (Option BAC  or BAH)
            Page
            Page
Chapter 2 397                
54. Comms Frequency Response (Option BAC or BAH)
              Table 2-108  Source/Splitter Characterization
              398 Chapter 2 
54. Comms Frequency Response (Option BAC or BAH)            
Page
            Page
              Flatness Error Mkr1 Amptd Current Channel B Splitter Tracking Error=
            Page
              Table 2-110  Frequency Response Worksheet Test:_______ Atten:_______ RBW:_______ Preamp:_______
            Page
            Page
              58. GSM Phase and Frequency Error (Options  BAH and B7E)
            Page
            Page
              59. Comms Absolute Power Accuracy (Options  BAC or BAH)
            Page
            Page
            Page
              Amplitude Accuracy   Measured Amplitude   Amptd 
              A
mplitude Accuracy Measured Amplitude Amptd15dBm ActualTotalAtten RefAtten0dB                
+()=
Measured Amplitude   Amptd  + ()                
<
              Table 2-120  GSM Transmit Power (Option BAH), Preamp Off, Input level 
Table 2-121  Amplitude Accuracy Worksheet, cdmaOne, Input Level <25  dBm                
Table 2-122  Amplitude Accuracy Worksheet, GSM, Input Level 20 dBm
            Page
                  Table 2-124  cdmaOne Channel Power Accuracy (PCS Band)
                  Table 2-125  GSM Transmit Power Accuracy (Cellular Band)
                    Table 2-126  GSM Transmit Power Accuracy (PCS Band)
                  RefAtten40dB 10 dB Actual40dB 1 dB Actual0dB
            Page
              Measured Amplitude   Amptd 
+() 40dB+                
Table 2-129  cdmaOne Receive Channel Power (Cellular Band) Preamp Off
                    Table 2-130  cdmaOne Receive Channel Power (PCS Band) Preamp Off
            Page
              RefAtten40dB 10 dB Actual40dB 1 dB Actual0dB
            Page
              Measured Amplitude   Amptd 
+() 40dB+                
 
              Table 2-134  cdmaOne Receive Channel Power (PCS Band) Preamp On
                     Table 2-133  cdmaOne Receive Channel Power (Cellular Band) Preamp On
            Page
              Agilent E4401B Performance Verification Test 
Tests for the Agilent E4401B only are included in this test record,               
Table 3-1  Agilent E4401B Performance Verification Test Record
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
              Table 3-2  Agilent E4401B Performance Verification Test Record
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
              Agilent E4402B Performance Verification Test 
Tests for the Agilent E4402B only are included in this test record,               
Table 3-3  Agilent E4402B Performance Verification Test Record
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
              30 40  60
  70 80             
Page
              Agilent E4403B Performance Verification 
Tests for the Agilent E4403B only are included in this test record,               
Table 3-5  Agilent E4403B Performance Verification Test Record
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
              Agilent E4404B Performance Verification Test 
Tests for the Agilent E4404B only are included in this test record,               
Table 3-7  Agilent E4404B Performance Verification Test Record
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
              Agilent E4405B Performance Verification 
Tests for the Agilent E4405B only are included in this test record,               
Table 3-9  Agilent E4405B Performance Verification Test Record
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
              Agilent E4407B Performance Verification 
Tests for the Agilent E4407B only are included in this test record,               
Table 3-11  E4407B Performance Verification Test Record
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
              Agilent E4408B Performance Verification Test 
Tests for the Agilent E4408B only are included in this test record,               
Table 3-13  Agilent E4408B Performance Verification Test Record
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
              Agilent E4411B Performance Verification Test 
Tests for the Agilent E4411B only are included in this test record,               
Table 3-15  Agilent E4411B Performance Verification Test Record
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
            Page
              What Youll Find in This Chapter 
              Before You Call Agilent Technologies 
Check the Basics               
Read the Warranty 
Service Options                 
How to Call Agilent Technologies 
Chapter 4 639                
If You Have a Problem Before You Call Agilent Technologies
              Table 4-1  Agilent Technologies Sales and Service Offices
              How to Return Your Analyzer for Service 
Service Tag                 
Original Packaging 
            Page
              Other Packaging