Chapter 2 185
Performance Verification Tests
23. Frequency Response, Agilent E4402B and E4403B
13.Note the most positive number from the Flatness Relative to 50 MHz
(dc) column of Table 2-45 for frequencies between 1.7 GHz and
2.0 GHz. Record this number in Table 2-47 as the Maximum
Response for Band 0B.
14.Note the most negative number from the Flatness Relative to
50 MHz (dc) column of Table 2-45 for frequencies between 1.7 GHz
and 2.0 GHz. Record this number in Table 2-47 as the Minimum
Response for Band 0B.
15.For Band 0B in Table 2-47, subtract the Minimum Response value
from the Maximum Response value and record the result in the
Peak-to-Peak Response column.
16.Transfer the values entered in Table 2-47 to the test record in this
guide.
17.The frequency response test for analyzers without Option UKB is
now complete. The following steps apply only to analyzers having
Option UKB.
18.Record the most positive number from the Flatness Relative to 50
MHz (ac) column of Table 2-45 as the Maximum Response for Band 0
in Table 2-48.
19.Record the most negative number from the Flatness Relative to
50 MHz (ac) column of Table 2-45 as the Minimum Response for
Band 0 in Table 2-48.
20.For Band 0 in Table 2-48, subtract the Minimum Response value
from the Maximum Response value and record the result in the
Peak-to-Peak Response column.
21.Note the most positive number from the Flatness Relative to 50 MHz
(ac) column of Table 2-45 for frequencies between 800 MHz and
1.0 GHz. Record this number in Table 2-48 as the Maximum
Response for Band 0A.
22.Note the most negative number from the Flatness Relative to
50 MHz (ac) column of Table 2-45 for frequencies between 800 MHz
and 1.0 GHz. Record this number in Table 2-48 as the Minimum
Response for Band 0A.
23.For Band 0A in Table 2-48, subtract the Minimum Response value
from the Maximum Response value and record the result in the
Peak-to-Peak Response column.
24.Note the most positive number from the Flatness Relative to 50 MHz
(ac) column of Table 2-45 for frequencies between 1.7 GHz and
2.0 GHz. Record this number in Table 2-48 as the Maximum
Response for Band 0B.
25.Note the most negative number from the Flatness Relative to
50 MHz (ac) column of Table 2-45 for frequencies between 1.7 GHz