512 Chapter 3
Performance Verification Test Records
Agilent E4404B Performance Verification Test Record
Maximum Response (19)________ 1.5 dB ±0.39 dB
Minimum Response –1.5 dB (20)________ ±0.39 dB
Peak-to-Peak Response (21)________ 2.6 dB ±0.39 dB
0 to 55° C
Non-Option UKB
dc Coupled
Band 0
(9 kHz to 3.0 GHz)
Maximum Response (1)________ 0.76 dB ±0.19 dB
Minimum Response –0.76 dB (2)________ ±0.19 dB
Peak-to-Peak Response (3)________ 1.52 dB ±0.19 dB
Option UKB
dc Coupled
Band 0
(100 Hz to 3.0 GHz)
Maximum Response (1)________ 1.0 dB ±0.19 dB
Minimum Response –1.0 dB (2)________ ±0.19 dB
Peak-to-Peak Response (3)________ 2.0 dB ±0.19 dB
dc Coupled
Band 1
(3.0 GHz to 6.7 GHz)
Maximum Response (4)________ 2.5 dB ±0.39 dB
Minimum Response –2.5 dB (5)________ ±0.39 dB
Peak-to-Peak Response (6)________ 3.0 dB ±0.39 dB
Table 3-8 Agilent E4404B Performance Verification Test Record
Agilent Technologies
Model E4404B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum Results
Measured Maximum Measurement
Uncertainty