Chapter 3 619
Performance Verification Test Records
Agilent E4411B Performance Verification Test Record
1490 MHz 1 MHz 1489.999998
MHz
(5)________ 1490.000002
MHz
±0 Hz
5. Frequency Span
Accuracy
Span Start Freq
1500 MHz 0 Hz 1185 MHz (1)________ 1215 MHz ±3.06 MHz
100 MHz 10 MHz 79 MHz (2)________ 81 MHz ±204 kHz
100 kHz 10 MHz 79 kHz (3)________ 81 kHz ±204 Hz
100 MHz 800 MHz 79 MHz (4)________ 81 MHz ±204 kHz
100 kHz 800 MHz 79 kHz (5)________ 81 kHz ±204 Hz
100 MHz 1400 MHz 79 MHz (6)________ 81 MHz ±204 kHz
100 kHz 1499 MHz 79 kHz (7)________ 81 kHz ±204 Hz
7. Noise Sidebands
Offset from 1 GHz signal
10 kHz (1)________ –93 dBc/Hz ±2.44 dB
20 kHz (2)________ –100 dBc/Hz ±2.44 dB
30 kHz (3)________ –104 dBc/Hz ±2.44 dB
100 kHz (4)________ –113 dBc/Hz ±2.44 dB
9. System-Related
Sidebands
Offset from 500 MHz
signal
30 kHz to 230 kHz (1)________ –65 dBc ±1.30 dB
–30 kHz to –230 kHz (2)________ –65 dBc ±1.30 dB
10. Residual FM
1 kHz Res BW (1)________ 150 Hz ±21 Hz
11. Sweep Time Accuracy
Table 3-16 Agilent E4411B Performance Verification Test Record
Agilent Technologies
Model E4411B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum Results
Measured Maximum Measurement
Uncertainty