Chapter 2 85
Performance Verification Tests
11. Sweep Time Accuracy
accuracy as follows:
NOTE The sweep time accuracy is defined as a percentage of the indicated
sweep time, not of the indicated signal separation. Therefore, it is
appropriate to divide the difference between the marker delta ( Mkr1)
reading and the nominal signal separation by the sweep time, rather
than dividing by the nominal signal separation.
12.Record the calculated sweeptime accuracy in Table 2-12.
13.If the analyzer is not equipped with Option AYX, fast time domain
sweeps, or Option B7D, DSP and Fast ADC, repeat step 9 through
step 12 only for sweeptime settings between 5 ms and 10 s as
indicated in Table 2-12. For each sweeptime setting, set the function
generator to the frequency indicated in Table 2-12.
14.If the analyzer is equipped with Option AYX, fast time domain
sweeps, or Option B7D, DSP and Fast ADC, repeat step 9 through
step 12 for all sweeptime settings as indicated in Table 2-12. For
each sweeptime setting, set the function generator to the frequency
indicated in Table 2-12.

Sweep Time Accuracy 100 Mkr1 0.8 Sweep Time×()

Sweep Time

------------------------------------------------------------------------

×=
Table 2-12 Sweep Time Accuracy
Analyzer Sweep
Time Setting
Synthesizer Function
Generator Frequency
Marker
Delta
( Mkr1)
Reading
Sweep
Time
Accuracy
(%)
Test
Record
Entry
5 ms 2.0 kHz 1)
20 ms 500.0 Hz 2)
100 ms 100.0 Hz 3)
1 s 10.0 Hz 4)
10 s 1.0 Hz 5)
The following entries only apply to analyzers equipped with Option AYX or B7D.
1 ms 10.0 kHz 6)
500 µs 20.0 kHz 7)
100 µs 100.0 kHz 8)