Chapter 3 553
Performance Verification Test Records
Agilent E4405B Performance Verification Test Record
9 kHz to 2.9 GHz (1)________ –16 dBm ±2.58 dB
2.9 GHz to 3.0 GHz (2)________ –16 dBm ±3.03 dB
50. Gate Delay Accuracy
and Gate Length
Accuracy
(Option 1D6 only)
Minimum Gate Delay 499.9 ns (1)________ 1.5001µs±549 ps
Maximum Gate Delay 499.9 ns (2)________ 1.5001µs±549 ps
1 µs Gate Length 499.9 ns (3)________ 1.5001µs±520 ps
65 ms Gate Length 64.993 ms (4)________ 65.007 ms ±647 ns
51. Gate Mode Additional
Amplitude Error (Option
1D6 only)
Amplitude Error –0.2 dB (1)________ 0.2 dB ±0.023 dB
58. GSM Phase and
Frequency Error
(Option BAH and B7E)
Peak Phase Error –2.1 Deg (1)________ 2.1 Deg 0.52 Deg
RMS Phase Error –1.1 Deg (2)________ 1.1 Deg 0.000063 Deg
Frequency Error –10 Hz (3)________ 10 Hz ±0.006 Hz
59. Comms Absolute Power
Accuracy
(Options BAC or BAH)
20 to 30° C
cdmaOne Channel Power
Accuracy
(Option BAC only)
Cellular Band
Input Amplitude
15 dBm –0.82 dB (1)________ 0.82 dB ±0.173 dB
Table 3-10 Agilent E4405B Performance Verification Test Record
Agilent Technologies
Model E4405B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum Results
Measured Maximum Measurement
Uncertainty