52 Chapter 2 
Performance Verification Tests
4. Frequency Readout and Marker Frequency Count Accuracy: Agilent E4404B, E4405B, 
E4407B, and E4408B
Procedure 
This performance verification test consists of two parts:
“Part 1: Frequency Readout Accuracy”
“Part 2: Marker Count Accuracy”
Perform “Part 1: Frequency Readout Accuracy” before “Part 2: Marker 
Count Accuracy.”
Part 1: Frequency Readout Accuracy 
1. Connect the equipment as shown in Figure 2-4. Remember to 
connect the 10 MHz REF OUT of the analyzer to the 10 MHz REF 
INPUT of the synthesized sweeper.
Option BAB only: Use the APC 3.5 adapter to connect the cable to 
the analyzer input.
2. Perform the following steps to set up the equipment:
a. Press INSTRUMENT PRESET on the synthesized sweeper, then set 
the controls as follows:
CW, 1.5 GHz
POWER LEVEL, –10 dBm
b. Press Preset on the analyzer. Press the Factory Preset softkey, if it 
is displayed. Set the analyzer by pressing the following keys:
FREQUENCY, 1.5 GHz
SPAN, 20 MHz
3. Press Peak Search (or Search) on the analyzer to measure the 
frequency readout accuracy.
4. Record the marker (Mkr1) frequency reading in the performance 
verification test record as indicated in Table 2-6.
5. Change to the next analyzer span setting listed in 
Table 2-6.
6. Repeat step 3 through step 5 for each analyzer frequency and span 
setting and synthesized sweeper CW frequency setting listed in 
Table 2-6 for the analyzer being tested.
Table 2-6  Frequency Readout Accuracy 
Synthesized 
Sweeper CW 
Frequency
(MHz)
Analyzer 
Span 
(MHz)
Analyzer 
Center 
Frequency
(GHz) 
Test 
Record 
Entry 
Frequency 
(GHz) 
1500 20 1.5 1)