544 Chapter 3
Performance Verification Test Records
Agilent E4405B Performance Verification Test Record
Option UKB
dc Coupled
Band 0
(100 Hz to 3.0 GHz)
Maximum Response (1)________ 1.0 dB ±0.19 dB
Minimum Response –1.0 dB (2)________ ±0.19 dB
Peak-to-Peak Response (3)________ 2.0 dB ±0.19 dB
dc Coupled
Band 1
(3.0 GHz to 6.7 GHz)
Maximum Response (4)________ 2.5 dB ±0.39 dB
Minimum Response –2.5 dB (5)________ ±0.39 dB
Peak-to-Peak Response (6)________ 3.0 dB ±0.39 dB
dc Coupled
Band 2
(6.7 GHz to 13.2 GHz)
Maximum Response (7)________ 3.0 dB ±0.40 dB
Minimum Response –3.0 dB (8)________ ±0.40 dB
Peak-to-Peak Response (9)________ 4.0 dB ±0.40 dB
ac Coupled
Band 0
(100 kHz to 3.0 GHz)
Maximum Response (10)________ 1.0 dB ±0.11 dB
Minimum Response –1.0 dB (11)________ ±0.11 dB
Peak-to-Peak Response (12)________ 2.0 dB ±0.11 dB
Table 3-10 Agilent E4405B Performance Verification Test Record
Agilent Technologies
Model E4405B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum Results
Measured Maximum Measurement
Uncertainty