570 Chapter 3
Performance Verification Test Records
Agilent E4407B Performance Verification Test Record
15. Reference Level
Accuracy
Log, Res BW 1 kHz
Reference Level
–10 dBm –0.3 dB (1)________ 0.3 dB ±0.091 dB
0 dBm –0.3 dB (2)________ 0.3 dB ±0.100 dB
–30 dBm –0.3 dB (3)________ 0.3 dB ±0.092 dB
–40 dBm –0.3 dB (4)________ 0.3 dB ±0.094 dB
–50 dBm –0.5 dB (5)________ 0.5 dB ±0.095 dB
–60 dBm –0.5 dB (6)________ 0.5 dB ±0.097 dB
–70 dBm –0.5 dB (7)________ 0.5 dB ±0.099 dB
–80 dBm –0.7 dB (8)________ 0.7 dB ±0.101 dB
Linear, Res BW 1 kHz
Reference Level
–10 dBm –0.3 dB (9)________ 0.3 dB ±0.091 dB
0 dBm –0.3 dB (10)________ 0.3 dB ±0.100 dB
–30 dBm –0.3 dB (11)________ 0.3 dB ±0.092 dB
–40 dBm –0.3 dB (12)________ 0.3 dB ±0.094 dB
–50 dBm –0.5 dB (13)________ 0.5 dB ±0.095 dB
–60 dBm –0.5 dB (14)________ 0.5 dB ±0.097 dB
–70 dBm –0.5 dB (15)________ 0.5 dB ±0.099 dB
–80 dBm –0.7 dB (16)________ 0.7 dB ±0.101 dB
Log, Res BW 300 Hz
(Option 1DR only)
Reference Level
–10 dBm –0.3 dB (17)________ 0.3 dB ±0.091 dB
Table 3-12 Agilent E4407B Performance Verification Test Record
Agilent Technologies
Model E4407B Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum Results
Measured Maximum Measurement
Uncertainty