32. Spurious Responses: Agilent E4404B, E4405B, E4407B, and E4408B
1.8 GHz low pass filter (2 required)
4.4 GHz low pass filter (2 required)
Cable, BNC, 120-cm (48-in) 
Cable, APC 3.5, 91-cm (48-in) (2 required)
Adapter, Type-N (m) to APC 3.5 (f) (3 required)
Adapter, Type-N (m) to SMA (m) 
Adapter, Type-N (m) to BNC (f) 
Adapter, Type-N (m) to Type-N (m) 
Adapter, SMA (f) to BNC (m) 
Adapter, APC 3.5 (f) to APC 3.5 (f) 
Procedure 
This performance test consists of two parts: 
Part 1: Third Order Intermodulation Distortion
Part 2: Second Harmonic Distortion
Perform Part 1 before Part 2. 
Part 1: Third Order Intermodulation Distortion  
1. Zero and calibrate the power meter and microwave power sensor in 
log mode (power reads out in dBm), as described in the power meter 
operation manual. 
2. Connect the equipment as shown in Figure 2-49 using the 300 MHz 
low pass filter with the output of the directional bridge connected to 
the power sensor.