54 Chapter 2
Performance Verification Tests
4. Frequency Readout and Marker Frequency Count Accuracy: Agilent E4404B, E4405B,
E4407B, and E4408B
verification test record as indicated in Table 2-7.
4. Repeat step 2 and step 3 for each analyzer center frequency and
span setting and synthesized sweeper CW frequency setting listed in
Table 2-7 for the analyzer being tested.
Performance verification test “Frequency Readout Accuracy and
Marker Count Accuracy” is now complete.
Table 2-7 Marker Count Accuracy
Synthesized Sweeper
CW Frequency
Analyzer
Center
Frequency
Analyzer
Span
Counter
Frequency
(Cntr1)
MHz GHz MHz Test Record
Entry
1500 1.5 20 16)
1500 1.5 1 17)
4000 4.0 20 18)
4000 4.0 1 19)
Stop here for Agilent E4404B.
9000 9.0 20 20)
9000 9.0 1 21)
Stop here for Agilent E4405B.
16000 16.0 20 22)
16000 16.0 1 23)
21000 21.0 20 24)
21000 21.0 1 25)