Chapter 2 87
Performance Verification Tests
12. Display Scale Fidelity
Figure 2-12 Scale Fidelity Test Setup
CAUTION Use only 75 cables, connectors, and adapters on instruments with
75 connectors, or the connectors will be damaged.
Procedure
Calculate the Actual Attenuation Errors
1. From the calibration data supplied with the 10 dB step attenuator,
enter into Column 4 of Table 2-13 through Table 2-17 the actual
attenuation for the corresponding nominal attenuation settings.
Enter data in Table 2-14, Table 2-16, and Table 2-17 only if the
analyzer has Option 1DR.
If the calibration data does not indicate an actual attenuation value
for the 0 dB setting, enter 0 dB.
NOTE The Agilent 8496G programmable attenuator has four attenuator
sections consisting of 10 dB, 20 dB, and 40 dB attenuators. If using the
Agilent 8496G programmable attenuator, enter the calibration data for
the section three, 40 dB step, rather than the section four, 40 dB step.
2. From the calibration data supplied with the 1 dB step attenuator,
enter into Column 5 of Table 2-13 through Table 2-17 the actual
attenuation for the corresponding nominal attenuation settings.
Enter data in Table 2-14, Table 2-16, and Table 2-17 only if the
analyzer has Option 1DR.