(023-100 intentionally excluded)

101 EIGS/MIGS Board Test

1Select: 101

2Press: READY to run the test.

3If an error is found:

The error code appears.

Look up the code in “Error Code/TAG Cross-Reference” on page 2-3to determine which TAG to follow.

Power-on-reset to exit.

4If no errors are found:

The test continues to run.

After about five minutes, press STOP to exit.

102 EIGS/MIGS Board Test (Continuous Loop)

1Go to: 102

2Press: READY to run the test.

3If an error is found:

The error code is displayed for at least 1 second.

Look up the code in “Error Code/TAG Cross-Reference” on page 2-3to determine which TAG to follow.

If the test resumes, press STOP to exit; if the test does not resume, power-on-reset.

4If no errors are found:

The test continues to loop.

Press STOP to exit.

5-20

Diagnostic Tests