3Z4L Laser Micrometer Protocol

Appendix K

 

 

 

Calibration Set (Sequence No. 404)

This sequence calibrates the Laser Micrometer.

Send Data Word Allocation (2nd Operand of PMCR)

First word of send +0

data

+1

 

 

+2

 

+3

 

+4

 

+5

 

+6

Number of send data words

Decimal portion

(Undefined)

Decimal integer portion

 

 

 

 

 

Sign

 

(Undefined)

 

 

 

 

 

 

 

 

Decimal portion

 

 

 

 

 

 

 

(Undefined)

(Decimal integer portion)

 

 

 

 

 

Sign

 

(Undefined)

 

 

 

 

 

 

 

High calibration gauge dimension

Low calibration gauge dimension

Offset

Contents (data format)

 

Data

 

 

 

 

 

 

 

 

 

+0

Number of send data words

0007 (fixed)

 

 

 

 

 

(4 digits BCD)

 

 

 

 

 

 

 

 

 

 

+1

HC gauge dimension

0000 to 9999

 

Example –123.4567

 

(Decimal portion)

 

 

 

 

 

 

(4 digits BCD)

 

 

+1

4 5 6 7

 

 

 

 

 

+2

0 1 2 3

 

 

 

 

 

 

 

 

 

 

 

 

+3

2 D 0 0

 

 

 

 

 

 

 

 

 

 

 

 

 

+2

HC gauge dimension

000 to 999

 

Example –123.4567

 

(integer portion)

 

 

 

 

 

 

 

 

+1

4 5 6 7

 

 

(3 digits BCD)

 

 

 

 

 

 

 

 

 

 

 

 

 

+2

0 1 2 3

 

+3

HC gauge dimension (Sign)

if +: 20(‘ ’)

 

 

 

 

 

 

 

+3

2 D 0 0

 

 

(ASCII 1 character)

if –: 2D(‘–’)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

+4 to

LC gauge dimensions

Same as HC gauge dimensions

 

 

+6

 

 

 

 

 

 

 

 

 

 

 

 

 

Note 1. For this sequence, both the high and low calibration gauges must be set.

2.The limit value, reference value, and offset value can be set to 3 digits for the integer portion and to 4 dig- its for the decimal portion.

Receive Data Word Allocation (3rd Operand of PMCR)

None.

Calibration Release (Sequence No. 405)

This sequence releases the calibration of the Laser Micrometer.

Send Data Word Allocation (2nd Operand of PMCR)

None.

Receive Data Word Allocation (3rd Operand of PMCR)

None.

Note This sequence releases both the high and low calibration.