3Z4L Laser Micrometer Protocol

Appendix K

 

 

 

Receive Data Word Allocation (3rd Operand of PMCR)

Receive data

+0

Number of receive data words

storage words

 

 

 

 

+1

(Undefined)

 

The number of measuring times

 

 

 

 

 

 

 

 

+2

 

 

Invalid time

 

 

 

 

 

 

+3

 

 

Decimal portion

 

 

 

 

 

 

+4

(Undefined)

 

Integer portion

 

 

 

 

 

 

+5

Sign

 

(Undefined)

 

 

 

 

 

 

+6

 

 

Decimal portion

 

 

 

 

 

 

+7

(Undefined)

 

Integer portion

 

 

 

 

 

 

+8

Sign

 

(Undefined)

 

 

 

 

 

Detection lower limit

Detection upper limit

Offset

Contents (data format)

Data

 

 

 

 

 

 

 

 

 

 

+0

Number of receive data

0009 (fixed)

 

 

 

 

words (4 digits BCD)

 

 

 

 

 

 

 

 

 

 

 

 

+1

The number of measuring

000 to 999

 

 

 

 

times

 

 

 

 

 

 

(3 digits BCD)

 

 

 

 

 

 

 

 

 

 

 

 

+2

Invalid time

0001 to 9999

 

 

 

 

(4 digits BCD)

 

 

 

 

 

 

 

 

 

 

 

+3

Detection lower limit

0000 to 9999

 

Example –123.4567

 

(decimal portion)

 

 

 

 

 

 

 

 

+3

4 5 6 7

 

 

(4 digits BCD)

 

 

 

 

 

 

 

 

 

 

 

 

 

+4

0 1 2 3

 

+4

Detection lower limit

000 to 999

 

 

 

(integer portion)

 

 

 

 

 

 

 

 

+5

F 0 0 0

 

 

(3 digits BCD)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

+5

Detection lower limit (Sign)

If +: 0

 

 

 

 

 

(BIN)

If –: F

 

 

 

 

 

 

 

 

 

 

 

+6 to

Detection upper limit

Same as detection lower limit

 

 

 

+8

 

 

 

 

 

 

 

 

 

 

 

 

 

Note This sequence cannot be used when DIP switch SW3, pin 5 of the Laser Micrometer is not turned ON.

3Z4L Initialize (3000-series) (Sequence No. 430)

This sequence clears the 3Z4L, sets the mm unit, sets memory switches, does not process statistics, and clears the statistic memory.

Send Data Word Allocation (2nd Operand of PMCR)

The send data word allocation is similar to that of sequence No. 401 (Memory Switch Setting)

Receive Data Word Allocation (3rd Operand of PMCR)

None.

339

Page 343
Image 343
Omron CQM1H-SCB41 operation manual 3Z4L Initialize 3000-series Sequence No