3Z4L Laser Micrometer Protocol

 

 

 

 

Appendix K

 

 

 

 

 

 

 

 

 

 

 

 

 

Offset

Contents (data format)

Data

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

+0

Number of receive data

0019 (fixed)

 

 

 

 

 

 

 

words (4 digits BCD)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

+1

Segment number

1 to 5

 

 

 

 

 

 

 

(1 digit BCD)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

+2

Measurement interval

1 to 7

 

 

 

 

 

 

 

number (1 digit BCD)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

+3

Lower limit value (Decimal

0000 to 9999

 

Example –123.4567

 

 

 

 

portion)

 

 

 

 

 

 

 

 

 

 

 

+3

4 5 6 7

 

 

 

 

 

(4 digits BCD)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

+4

0 1 2 3

 

 

 

 

+4

Lower limit value (Integer

000 to 999

 

 

 

 

 

 

portion)

 

 

 

 

 

 

 

 

 

 

 

+5

F 0 0 0

 

 

 

 

 

(3 digits BCD)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

+5

Lower limit value (Sign)

If +: 0

 

 

 

 

 

 

 

 

(BIN)

If –: F

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

+6 to

Upper limit value

Same as lower limit value

 

 

 

 

 

+8

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

+9 to

Reference value

Same as lower limit value

 

 

 

 

 

+11

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

+12

Analog output scale number

0 to 3

 

 

 

 

 

 

 

(1 digit BCD)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

+13 to

Forced zero number (ASCII

5A45524F2B (“ZERO+”)

 

 

 

 

 

+15

5 characters)

4E4F524D20 (“NORM ”)

 

 

 

 

 

 

 

 

5A45524F2D (“ZERO–”)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

+16

Data output condition

0 to 6

 

 

 

 

 

 

 

(1 digit BCD)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

+17

Data output timer value

000 to 999

 

 

 

 

 

 

 

(3 digits BCD)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

+18

Number of seconds for latch

00 to 99

 

 

 

 

 

 

 

timer

 

 

 

 

 

 

 

 

 

(2 digits BCD)

 

 

 

 

 

 

Note This sequence can be used for the 3Z4L-4000 Series only when pin 8 on DIP switch SW2 is turned ON.

Single Run Measurement Start (4000-series) (Sequence No. 434)

When the sample measurement condition is from 1 to 999, this sequence performs a single run measurement and requests the measurement results.

Send Data Word Allocation (2nd Operand of PMCR)

None.

Receive Data Word Allocation (3rd Operand of PMCR)

Receive data

+0

Number of receive data words

storage words

 

 

 

+1

 

Decision result

 

 

 

 

 

 

 

+2

 

Decimal portion

 

 

 

 

 

+3

(Undefined)

Integer portion

 

+4

Sign

(Undefined)

 

+5

 

Decimal portion

 

 

 

 

+6

(Undefined)

Integer portion

 

 

 

 

+7

Sign

(Undefined)

 

 

 

Measurement value

Deviation value

343

Page 347
Image 347