14. Reference Level Accuracy, HP 8591EM
4MKR5 More 1 of 3 MK TRACK ON OFF (OFF)
13.Set the synthesizer/level generator amplitude and EMC analyzer reference level according to Table
14.Record the MKR 1 amplitude reading in Table
Table
Synthesizer/Level | EMC Analyzer | MKR 1 Reading (dB) | ||
Generator Amplitude | Reference Level |
|
|
|
|
|
|
|
|
(dBm) | (dBV) | Min. | TR Entry | Max. |
|
|
|
|
|
010 | 87 | 0 (Ref) | 0 (Ref) | 0 (Ref) |
0 | 97 | 00.4 | 10 | +0.4 |
+10 | 107 | 00.5 | 11 | +0.5 |
020 | 77 | 00.4 | 12 | +0.4 |
030 | 67 | 00.5 | 13 | +0.5 |
040 | 57 | 00.8 | 14 | +0.8 |
050 | 47 | 01.0 | 15 | +1.0 |
060 | 37 | 01.1 | 16 | +1.1 |
070 | 27 | 01.2 | 17 | +1.2 |
080 | 17 | 01.3 | 18 | +1.3 |
|
|
|
|
|
Narrow Bandwidths
15. Press the following EMC analyzer keys:
4AMPLITUDE5 87 4+dBV5 SCALE LOG LIN (LOG) 1 4dB5
4BW5 IF BW AUTO MAN (MAN) 300 4Hz5
4SPAN5 10 4kHz5
4SWEEP/TRIG5 SWEEP CONT SGL (CONT)
16.Set the synthesizer/level generator to 010 dBm.
17.Repeat steps 4 through 6, using Table
Performance Verication Tests