30. Spurious Response, HP 8593EM

4AMPLITUDE5 REF LVL 97 dB￿V

4MKR !5 MARKER ! HIGH More 1 of 3 PEAK EXCURSN 3 4dB5 4DISPLAY5 More 1 of 2 THRESHLD ON OFF (ON) 17 dB￿V

38.On synthesized sweeper #1, set RF on. Adjust the power level until the measuring receiver reads 012 dBm 60.05 dB.

39.Disconnect the power sensor from the directional coupler. Connect the directional coupler directly to the EMC analyzer INPUT 50 ￿ using an adapter (do not use a cable).

Option 026 only: Connect the directional coupler directly to the EMC analyzer INPUT 50 ￿.

40. On the EMC analyzer, press the following keys:

4MKR !5 MARKER ! HIGH

4MKR5 More 1 of 3 MK TRACK ON OFF (ON)

4SPAN5 200 4kHz5

Wait for the AUTO ZOOM message to disappear.

4MKR5 More 1 of 3 MK TRACK ON OFF (OFF) 4FREQUENCY5 4 * 5 (step-up key)

4MKR !5 MARKER ! HIGH

More 1 of 3 MARKER ! REF LVL

41.On synthesized sweeper #2, set RF on. Adjust the power level until the two signals are displayed at the same amplitude.

If necessary, adjust the EMC analyzer center frequency until the two signals are centered on the display.

42.Set the EMC analyzer by pressing the following keys:

4BW5 IF BW AUTO MAN (MAN) 1 4kHz5

AVG BW AUTO MAN (MAN) 100 4Hz5

2-196 Performance Veri￿cation Tests