13. Scale Fidelity, HP 8590 EM-Series

Table 2-9. Scale Fidelity, Linear Mode

Synthesizer/Level

% of

 

MKR Reading

 

Generator

Ref Level

 

 

 

 

 

 

 

 

Min. (mV)

TR Entry

Max. (mV)

 

Nominal Amplitude

(nominal)

 

 

 

 

 

 

 

 

 

 

 

+10 dBm

100

 

0 (Ref)

 

0 (Ref)

 

0 (Ref)

 

+7 dBm

70.7

151.59

65

165.01

 

+4 dBm

50

105.36

66

118.78

 

+1 dBm

35.48

72.63

67

86.05

 

02 dBm

25

49.46

68

62.88

 

 

 

 

 

 

 

 

 

 

24. Press the following EMC analyzer keys:

4BW5 IF BW AUTO MAN (MAN) 300 4Hz5

4SPAN5 10 4kHz5

25.Repeat steps 17 through 22 for the narrow IF bandwidths. Record the results as indicated in Table 2-10.

The scale ￿delity in linear mode is complete. Continue with step 26.

Table 2-10.

Scale Fidelity, Linear Mode for Narrow Bandwidths

Synthesizer/Level

% of

 

MKR Reading

Generator

Ref Level

 

 

 

 

 

 

 

 

Min. (mV)

 

 

TR Entry

 

Max. (mV)

Nominal Amplitude

(nominal)

 

 

 

 

 

 

 

 

 

 

 

 

 

+10 dBm

100

 

0 (Ref)

 

 

0 (Ref)

 

0 (Ref)

+7 dBm

70.7

151.59

 

69

 

165.01

+4 dBm

50

105.36

 

70

 

118.78

+1 dBm

35.48

72.63

 

71

 

86.05

02 dBm

25

49.46

 

72

 

82.88

 

 

 

 

 

 

 

 

 

Performance Veri￿cation Tests 2-67