19. Frequency Response, HP 8591EM
22.Set the EMC analyzer and the synthesizer/level generator to the next frequency settings listed in Table
23.At each frequency, adjust the synthesizer/level generator amplitude for a
24.Record the synthesizer/level generator amplitude setting in column 2 of Table
25.Repeat steps 22 through 24 for each frequency setting listed in Table
26.For each of the frequencies in Table
27.Add to each of the Response Relative to 50 MHz entries in Table
Test Results
Perform the following steps to verify the frequency response of the EMC analyzer.
1.Enter the most positive number from Table
2.Enter the most positive number from Table
3.Record the more positive of numbers from steps 1 and 2 in TR Entry 1 of the performance verication test record.
4.Enter the most negative number from Table
dB
5.Enter the most negative number from Table
6.Record the more negative of numbers from steps 4 and 5 in TR Entry 2 of the performance verication test record.
7.Subtract the results of step 6 from the results of step 3. Record this value in TR Entry 3 of the performance verication test record.