Measuring Fundamental Amplitudes
7.Set the EMC analyzer center frequency to the Fundamental Frequency listed in Table
8.Set the microwave spectrum analyzer controls as follows:
SPAN :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: 100 kHz
REFERENCE LEVEL :: :: :: :: ::: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: : +5 dBm
ATTEN :: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: : 20 dB
9.Set the microwave spectrum analyzer CENTER FREQUENCY to the Fundamental Frequency listed in Table
10.On the microwave spectrum analyzer, press PEAK SEARCH. Press MARKER
!REF LVL. Wait for another sweep to nish.
11.Record the microwave spectrum analyzer marker amplitude reading in Table
12.Repeat steps 8 through 11 for all Fundamental Frequency settings in Table
Table
Fundamental Fundamental
Frequency Amplitude
(dBm)
10MHz
900MHz
1.8 GHz
Measuring
13.On the EMC analyzer, set the center frequency to 10 MHz.
14.Set the microwave spectrum analyzer START FREQ, STOP FREQ, and RES BW as indicated in the rst row of Table
15.Press SINGLE on the microwave spectrum analyzer and wait for the sweep to nish. Press 4MKR !5 MKR ! HIGH .