33. Spurious Response, HP 8596EM

50.On the EMC analyzer, press 4PRESET5, then wait until the preset routine is ￿nished. Set the EMC analyzer by pressing the following keys:

4FREQUENCY5 4.0 4GHz5

4SPAN5 1 4MHz5

4AMPLITUDE5 REF LVL 97 4+dB￿V5

4MKR !5 MARKER ! HIGH More 1 of 3 PEAK EXCURSN 3 4dB5 4DISPLAY5 More 1 of 2 THRESHLD ON OFF 17 4+dB￿V5

51.On synthesized sweeper #1, set RF on. Adjust the power level until the measuring receiver reads 012 dBm 60.05 dB.

52.Disconnect the power sensor from the directional coupler. Connect the directional coupler directly to the EMC analyzer INPUT 50 ￿ using an adapter (do not use a cable).

53.On the EMC analyzer, press the following key:

4MKR !5 MARKER ! HIGH

4AMPLITUDE5 More 1 of 3 PRESEL PEAK

Wait for the peaking message to disappear.

4MKR5 More 1 of 3 MK TRACK ON OFF (ON)

4SPAN5 200 4kHz5

Wait for the AUTO ZOOM message to disappear, then press the following EMC analyzer keys:

4MKR5 More 1 of 3 MK TRACK ON OFF (OFF) 4FREQUENCY5 4 * 5 (step-up key)

4MKR !5 MARKER ! HIGH

More 1 of 3 MARKER ! REF LVL

54.On synthesized sweeper #2, set RF on. Adjust the power level until the two signals are displayed at the same amplitude.

If necessary, adjust the EMC analyzer center frequency until the two signals are centered on the display.

2-234 Performance Veri￿cation Tests