55.Harmonic Spurious Outputs, HP 8593EM, HP 8594EM, HP 8595EM, and HP 8596EM Option 010

b.Press 42 0 22 GHz5 (INSTR PRESET), 4RECALL5, 8. Adjust AMPTD CAL for a marker amplitude reading of 010 dBm.

c.Press 4RECALL5, 9. Adjust FREQ ZERO for a maximum amplitude response.

d.Press 4SHIFT5, 4FREQUENCY SPAN5 to start the 30 second internal error correction routine.

e.When the CALIBRATING! message disappears, press 4SHIFT5, 4START FREQ5 to use the error correction factors just calculated.

2.Connect the Type N cable between the RF OUT 50 ￿ and INPUT 50 ￿ connectors on the EMC analyzer. See Figure 2-80.

Figure 2-80.

Harmonic Spurious Outputs Test Setup,

HP 8593EM, HP 8594EM, HP 8595EM, and HP 8596EM

3.Press 4PRESET5 on the EMC analyzer, then wait for the preset routine to ￿nish. Set the EMC analyzer by pressing the following keys:

4FREQUENCY5 Band Lock 0-2.9 Gz BAND 0 (Except HP 8594EM)

4FREQUENCY5 300 4MHz5

4SPAN5 0 4Hz5

4BW5 30 4kHz5

4MKR5

4AUX/USER5 Track Gen SRC PWR ON OFF (ON) 102 4dB￿V5 TRACKING PEAK

Performance Veri￿cation Tests 2-351