62.CISPR Pulse Response, HP 8590 EM-Series

82.Press the following EMC analyzer keys:

4SPAN5 ZERO SPAN

4SWEEP/TRIG5 SWP TIME AUTO MAN (MAN) 2 4SEC5

4DET5 QP ON OFF (ON)

Note that this routine will take approximately 1 minute to execute.

4MKR !5 MARKER ! HIGH

83.Record the marker amplitude reading in:

Table 2-81 as the Measured 100 Hz Amplitude for the 120 kHz EMI bandwidth.

Table 2-82 as the Measured Relative Equivalent Level of Pulse for Bands C and D,

100 Hz Repetition Frequency.

84.Set PERIOD to 1 ms on the pulse generator.

85.On the EMC analyzer, press:

4SGL SWP5

4MKR !5 MARKER ! HIGH

86.Record the marker amplitude reading in Table 2-82 as the Measured Relative Equivalent Level of Pulse for Bands C and D, 1000 Hz Repetition Frequency.

87.Set PERIOD to .1 ms on the pulse generator.

88.On the EMC analyzer, press 4SGL SWP5 MARKER ! HIGH .

89.Record the marker reading in Table 2-83 as the Measured Relative Equivalent Level of Pulse for Bands C and D (QP value) for 10 kHz Repetition Frequency.

90.On the EMC analyzer, press the following keys:

4DET5 AVG ON OFF (ON)

4SGL SWP5

4MKR !5 MARKER ! HIGH

91.Record the marker amplitude reading in Table 2-83 as the Measured Relative Equivalent Level of Pulse for Band C/D (AVG) 10,000 Hz Repetition Frequency.

Performance Veri￿cation Tests 2-397