62.CISPR Pulse Response, HP 8590
82.Press the following EMC analyzer keys:
4SPAN5 ZERO SPAN
4SWEEP/TRIG5 SWP TIME AUTO MAN (MAN) 2 4SEC5
4DET5 QP ON OFF (ON)
Note that this routine will take approximately 1 minute to execute.
4MKR !5 MARKER ! HIGH
83.Record the marker amplitude reading in:
Table
Table
100 Hz Repetition Frequency.
84.Set PERIOD to 1 ms on the pulse generator.
85.On the EMC analyzer, press:
4SGL SWP5
4MKR !5 MARKER ! HIGH
86.Record the marker amplitude reading in Table
87.Set PERIOD to .1 ms on the pulse generator.
88.On the EMC analyzer, press 4SGL SWP5 MARKER ! HIGH .
89.Record the marker reading in Table
90.On the EMC analyzer, press the following keys:
4DET5 AVG ON OFF (ON)
4SGL SWP5
4MKR !5 MARKER ! HIGH
91.Record the marker amplitude reading in Table
Performance Verication Tests