4. Frequency Readout and Marker Count Accuracy, HP 8591EM and HP 8594EM
Part 1: Frequency Readout A ccuracy
1.Connect the equipment as shown in Figure
2.Perform the following steps to set up the equipment:
Press INSTRUMENT PRESET on the synthesized sweeper, then set the controls as follows:
CW :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: ::: :: :: :: :: : 1.5 GHz
POWER LEVEL ::: :: :: :: :: ::: :: :: :: ::: :: :: :: ::: :: :: :: :: :: 010 dBm
On the EMC analyzer, press 4PRESET5, wait for the preset routine to nish, then press 4FREQUENCY5 1.5 4GHz5.
3.Set the EMC analyzer to measure the frequency readout accuracy by pressing the following keys:
4SPAN5 20 4MHz5
4MKR !5 MARKER ! HIGH
4.Record the MKR frequency reading as TR Entry 1 in the performance verication test record. The reading should be within the limits shown in Table
5.Change to the next EMC analyzer span setting listed in Table
6.Repeat steps 3 through 5 for each EMC analyzer span setting/TR entry listed in Table
Table
EMC Analyzer | MKR Reading | ||
|
|
|
|
Span | Min. | TR Entry | Max. |
(MHz) | (MHz) |
| (MHz) |
|
|
|
|
20 | 1.49918 | 1 | 1.50082 |
10 | 1.49958 | 2 | 1.50042 |
1 | 1.499968 | 3 | 1.500032 |
|
|
|
|
7.Set the EMC analyzer by pressing the following keys:
4BW5 IF BW AUTO MAN (MAN) 300 4Hz5
Performance Verication Tests