33. Spurious Response, HP 8596EM
b.Press 4PRESET5 on the EMC analyzer, then wait for the preset routine to nish. Set the EMC analyzer by pressing the following keys:
4FREQUENCY5 5.6 4GHz5
4SPAN5 ZERO SPAN
4AMPLITUDE5 REF LVL 67 4+dBV5
4BW5 IF BW AUTO MAN (MAN) 1 4kHz5 AVG BW AUTO MAN (MAN) 30 4Hz5
4TRACE5 More 1 of 4 More 2 of 4 VID AVG ON OFF (ON) 10
4ENTER5
4SWEEP/TRIG5 SWP TIME AUTO MAN (MAN) 5.0 4sec5
c.Press 4SGL SWP5. Wait until AVG 10 is displayed along the left side of the CRT display.
d.Press 4MKR !5 MARKER ! HIGH on the EMC analyzer and record the marker amplitude reading as the Noise Level at 5.6 GHz in Table
12.Press 4PRESET5 on the EMC analyzer, then wait for the preset routine to nish. Set the EMC analyzer by pressing the following keys:
4FREQUENCY5 More 1 of 2 Band Lock
4FREQUENCY5 2.8 4GHz5
4SPAN5 10 4MHz5
13.Connect the equipment as shown in Figure
14.On the synthesized sweeper, press preset, then set the controls as follows:
CW :: :: :: ::: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: :: :: 2.8 GHz
POWER LEVEL : : : : : : : : : :: : : : : : : : : : : : : : : : : : : : : : : : :: : : : : : : : : : : 0 dBm
15. On the EMC analyzer, press the following keys:
4MKR !5 MARKER ! HIGH
4AMPLITUDE5 More 1 of 3 PRESEL PEAK
Wait for the peaking message to disappear.
Performance Verication Tests