4. Frequency Readout and Marker Count Accuracy, HP 8591EM and HP 8594EM
10. Set the EMC analyzer by pressing the following keys:
4BW5 IF BW AUTO MAN (MAN) 30 4Hz5
4SPAN5 2 4kHz5
11.Press 4MKR !5 MARKER ! HIGH 4MKR5 More 1 of 3 MK TRACK ON OFF
(ON), then wait until the count is completed (it may take several seconds).
12.Record the MKR reading as TR Entry 8 of the Performance verication Test Record. The reading should be within the limits of 1.49999989 and 1.50000011.