38. Gain Compression, HP 8596EM

35.On the EMC analyzer, press 4PRESET5, then wait for the preset routine to ￿nish. Press the EMC analyzer keys as follows:

4FREQUENCY5 50 4MHz5

4SPAN5 10 4MHz5

4AMPLITUDE5 97 4dB￿V5

4MKR !5 MKR ! HIGH

4MKR5 More 1 of 3 MK TRACK ON OFF (ON)

4SPAN5 2 4kHz5

Wait for the AUTO ZOOM message to disappear.

36.On synthesized sweeper #2, adjust the amplitude to place the signal 10 dB below the EMC analyzer reference level.

37.On the EMC analyzer, press 4SGL SWP5, then wait for the completion of a new sweep. Press 4MKR !5 MKR ! HIGH More 1 of 3 More 2 of 3 MARKER 1 .

38.On synthesized sweeper #1, set RF to ON.

39.On the EMC analyzer, press 4SGL SWP5, then wait for the completion of a new sweep. Press 4MKR !5 MKR ! HIGH .

40.Read the MKR 1 amplitude and record in the performance veri￿cation test record as TR Entry 3.

2-264 Performance Veri￿cation Tests