52.Tracking Generator Level Flatness, HP 8591EM Option 010 Table 2-65. Tracking Generator Level Flatness Worksheet

Center

 

Level

Cal

Freq

 

Flatness

Factor

 

 

(dB)

(MHz)

 

 

 

 

 

100 kHz

 

 

 

0.1

300 kHz

 

 

 

0.3

500 kHz

 

 

 

0.3

1 MHz

 

 

 

1

2 MHz

 

 

 

3

5 MHz

 

 

 

3

 

 

 

10 MHz

 

 

 

 

 

 

 

10

20 MHz

 

 

 

30

50 MHz

 

 

 

50

100 MHz

 

 

 

100

200 MHz

 

 

 

300

300 MHz

 

0 (Ref)

300

400 MHz

 

 

 

300

500 MHz

 

 

 

300

 

 

 

 

 

 

 

 

Center

 

Level

Cal

Freq

 

Flatness

Factor

 

 

 

(dB)

(MHz)

 

 

 

 

 

600 MHz

 

 

 

300

700 MHz

 

 

 

1000

800 MHz

 

 

 

1000

900 MHz

 

 

 

1000

1000

MHz

 

 

 

1000

1100

MHz

 

 

 

1000

 

 

 

1200

MHz

 

 

 

 

 

 

 

1000

1300

MHz

 

 

 

1000

1400

MHz

 

 

 

1000

1500

MHz

 

 

 

2000

1600

MHz

 

 

 

2000

1700

MHz

 

 

 

2000

1800

MHz

 

 

 

2000

 

 

 

 

 

 

13.Locate the most positive Level Flatness reading in Table 2-65 for the frequency ranges listed in Table 2-66 and record as the Maximum Flatness in the performance veri￿cation test record as shown in Table 2-66.

Table 2-66. Maximum Flatness

Description

TR Entry

 

(Maximum Flatness)

 

 

100 kHz

1

300 kHz to 5 MHz

2

10 MHz to 1800 MHz

3

 

 

Performance Veri￿cation Tests 2-337