16.Absolute Amplitude Calibration and Resolution (IF) Bandwidth Switching Uncertainties,HP 8590
This performance verication test must be performed with the EMC analyzer set in the spectrum analyzer mode.
To measure the absolute amplitude calibration uncertainty the input signal is measured after the
To measure the resolution (IF) bandwidth switching uncertainty an amplitude reference is taken with the bandwidth set to 3 kHz using the
Equipment Required
Cable, BNC, 23 cm (9 in)
Adapter, Type N (m) to BNC (f)
Additional Equipment for Option 026
Adapter, APC 3.5 (f) to Type N (f)
Figure
Performance Verication Tests