28. Other Input Related Spurious Responses, HP 8596EM

Figure 2-37.

Other Input Related Spurious Responses Test Setup, HP 8596EM

Procedure

Band 0

1.Zero and calibrate the measuring receiver and power sensor in log mode (power reads out in dBm), as described in the measuring receiver operation manual. Enter the power sensor 2 GHz Cal Factor into the measuring receiver.

2.Press INSTRUMENT PRESET on the synthesized sweeper and set the controls as follows:

CW : :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: ::: :: :: :: : 2000 MHz

POWER LEVEL :: :: ::: :: :: :: ::: :: :: :: :: ::: :: :: :: ::: :: :: :: ::: :: :: :: :: 04 dBm

3.Connect the equipment as shown in Figure 2-37. Connect the output of the synthesizer to the power sensor.

4.Press 4PRESET5 on the EMC analyzer, then wait for the preset routine to ￿nish. Set the EMC analyzer by pressing the following keys:

4FREQUENCY5 2.0 4GHz5

4SPAN5 1 4MHz5

4AMPLITUDE5 REF LVL 97 4dB￿V5

ATTEN AUTO MAN (MAN) 0 4dB5

Performance Veri￿cation Tests 2-175