19. Frequency Response, HP 8591EM
4MKR5 More 1 of 3 MK TRACK ON OFF (ON)
5.Adjust the synthesized sweeper power level for a
6.Set the sensor Cal Factor on the measuring receiver, then press RATIO.
7.Set the synthesized sweeper CW to 50 MHz.
8.Press 4FREQUENCY5 50 4MHz5 on the EMC analyzer.
9.Adjust the synthesized sweeper power level for an EMC analyzer
10.Set the sensor Cal Factor on the measuring receiver, then record the negative of the power ratio displayed on the measuring receiver in column 2 of Table
11.Set the synthesized sweeper CW to 100 MHz.
12.Press 4FREQUENCY5 100 4MHz5 on the EMC analyzer.
13.Adjust the synthesized sweeper power level for an EMC analyzer
14.Set the sensor Cal Factor on the measuring receiver, then record the negative of the power ratio displayed on the measuring receiver in column 2 of Table
15.On the synthesized sweeper, press CW, and 4 * 5
16.Record the negative of the power ratio displayed on the measuring receiver in column 2 of Table
17.Repeat steps 15 through 16 for each new frequency, entering the power sensor Cal Factor into the measuring receiver for each frequency setting as indicated in Table