19. Frequency Response, HP 8591EM

4MKR5 More 1 of 3 MK TRACK ON OFF (ON)

5.Adjust the synthesized sweeper power level for a MKR-TRK amplitude reading of 93 dB￿V 60.05 dB.

6.Set the sensor Cal Factor on the measuring receiver, then press RATIO.

7.Set the synthesized sweeper CW to 50 MHz.

8.Press 4FREQUENCY5 50 4MHz5 on the EMC analyzer.

9.Adjust the synthesized sweeper power level for an EMC analyzer MKR-TRK amplitude reading of 93 dB￿V 60.05 dB.

10.Set the sensor Cal Factor on the measuring receiver, then record the negative of the power ratio displayed on the measuring receiver in column 2 of Table 2-24 as the Error Relative to 300 MHz at 50 MHz.

11.Set the synthesized sweeper CW to 100 MHz.

12.Press 4FREQUENCY5 100 4MHz5 on the EMC analyzer.

13.Adjust the synthesized sweeper power level for an EMC analyzer MKR-TRK amplitude reading of 93 dB￿V 60.05 dB.

14.Set the sensor Cal Factor on the measuring receiver, then record the negative of the power ratio displayed on the measuring receiver in column 2 of Table 2-24 as the Error Relative to 300 MHz at 100 MHz.

15.On the synthesized sweeper, press CW, and 4 * 5 (step-up key), then on the EMC analyzer, press 4FREQUENCY5, and 4 * 5 (step-up key).

16.Record the negative of the power ratio displayed on the measuring receiver in column 2 of Table 2-24.

17.Repeat steps 15 through 16 for each new frequency, entering the power sensor Cal Factor into the measuring receiver for each frequency setting as indicated in Table 2-24.

2-100 Performance Veri￿cation Tests