39. Displayed Average Noise Level, HP 8591EM
1 MHz
10. Press the following EMC analyzer keys:
4FREQUENCY5 1 4MHz5
4SGL SWP5
Wait for the completion of a new sweep.
11. Press the following EMC analyzer keys:
4DISPLAY5 DSP LINE ON OFF (ON)
Adjust the display line so that it is centered on the average trace noise, ignoring any residual responses (refer to the Residual Responses verication test for any suspect residuals).
12.Record the display line amplitude setting as TR Entry 2 of the performance verication test record as the noise level at 1 MHz. The average noise level should be less than the specied limit.
1 MHz to 1.5 GHz
13. Press the following EMC analyzer keys:
4FREQUENCY5 START FREQ 1 4MHz5 STOP FREQ 1.5 4GHz5
4BW5 1 4MHz5
AVG BW AUTO MAN (MAN) 10 4kHz5
4SWEEP/TRIG5 SWEEP CONT SGL (CONT)
If the IF overload message is displayed on the EMC analyzer, performing the next step should clear the message.
14.Press 4FREQUENCY5 and adjust the center frequency setting, if necessary, to place the LO feedthrough just
15.Press the following EMC analyzer keys:
4SGL SWP5
4TRACE5 CLEAR WRITE A
More 1 of 4 More 2 of 4 VID AVG ON OFF (ON) 10 4Hz5
Wait until AVG 10 is displayed to the left of the graticule (the EMC analyzer will take ten sweeps, then stop).