20. Frequency Response, HP 8593EM
Frequency Response , Band 4
40. On the EMC analyzer, press the following keys:
4FREQUENCY5 More 1 of 2 Band Lock
4FREQUENCY5 19.1 4GHz5
CF STEP AUTO MAN (MAN) 100 4MHz5 (For Option 026 or 027 only, press:
200 4MHz5)
4SPAN5 5 4MHz5
4BW5 IF BW AUTO MAN (MAN) 1 4MHz5
AVG BW AUTO MAN (MAN) 10 4kHz5
4MKR !5 MARKER ! HIGH
4MKR5 More 1 of 3 MK TRACK ON OFF (ON)
41.Set the synthesized sweeper CW to 19.1 GHz.
42.On the EMC analyzer, press 4AMPLITUDE5, More 1 of 3 , and PRESEL PEAK .
43.Adjust the synthesized sweeper power level for an EMC analyzer
44.Record the negative of the power ratio displayed on the measuring receiver in Table
45.On the synthesized sweeper, press CW, and 4 * 5 (step up) key, then on the EMC analyzer, press 4FREQUENCY5, 4 * 5 (step up) key to step through the remaining frequencies listed in Table
46.At each new frequency repeat steps 42 through 44, entering the power sensor Cal Factor into the measuring receiver as indicated in Table
Performance Verication Tests