25.Other Input Related Spurious Responses, HP 8593EM
25.Other Input Related Spurious Responses ,HP 8593EMThis performance verication test must be performed with the EMC analyzer set in the spectrum analyzer mode.
A synthesized source and the EMC analyzer are set to the same frequency and the amplitude of the source is set to 0 dBm (107 dBV). A
There are no related adjustment procedures for this performance test.
Equipment Required
Synthesized sweeper
Measuring receiver (used as a power meter)
Power sensor, 50 MHz to 26.5 GHz
Power splitter
Adapter, Type N (m) to APC 3.5 (m)
Adapter, APC 3.5 (f) to APC 3.5 (f)
Cable, APC 3.5, 91 cm (36 in)
Performance Verication Tests