55.Harmonic Spurious Outputs, HP 8593EM, HP 8594EM, HP 8595EM, and HP 8596EM Option 010

the microwave spectrum analyzer frequency is the same as the Tracking Generator Output Frequency.

8.Locate the 2nd Harmonic Level for 9 kHz in Table 2-71 and record as TR Entry 1 of the performance veri￿cation record.

9.Locate the most positive 2nd Harmonic Level in Table 2-71 and record as TR Entry 2 of the performance veri￿cation test record.

10.Locate the 2nd Harmonic Level for 1.4 GHz in Table 2-71 and record as TR Entry 3 of the performance veri￿cation test record.

11.Locate the 3rd Harmonic Level for 9 kHz in Table 2-71 and record as TR Entry 4 of the performance veri￿cation record.

12.Locate the most positive 3rd Harmonic Level in Table 2-71 and record as TR Entry 5 of the performance veri￿cation test record.

Table 2-71. Harmonic Spurious Responses Worksheet

Tracking Generator

2nd Harmonic Level

3rd Harmonic Level

Frequency

(dBc)

(dBc)

 

 

 

9kHz

25 kHz

300kHz

100 MHz

300 MHz

900 MHz

1.4 GHz

 

N/A

Performance Veri￿cation Tests 2-353