55.Harmonic Spurious Outputs, HP 8593EM, HP 8594EM, HP 8595EM, and HP 8596EM Option 010
the microwave spectrum analyzer frequency is the same as the Tracking Generator Output Frequency.
8.Locate the 2nd Harmonic Level for 9 kHz in Table
9.Locate the most positive 2nd Harmonic Level in Table
10.Locate the 2nd Harmonic Level for 1.4 GHz in Table
11.Locate the 3rd Harmonic Level for 9 kHz in Table
12.Locate the most positive 3rd Harmonic Level in Table
Table
Tracking Generator | 2nd Harmonic Level | 3rd Harmonic Level |
Frequency | (dBc) | (dBc) |
|
|
|
9kHz
25 kHz
300kHz
100 MHz
300 MHz
900 MHz
1.4 GHz |
| N/A |
Performance Verication Tests