53. Tracking Generator Level Flatness,
HP 8593EM, HP 8594EM, HP 8595EM, HP 8596EM Option 010
Adapter, Type N (m) to BNC (f)
Figure
Tracking Generator Level Flatness Test Setup,
HP 8593EM, HP 8594EM, HP 8595EM, and HP 8596EM
Procedure
1.Connect the Type N cable between the RF OUT 50 and INPUT 50 connectors on the EMC analyzer. See Figure
2.Press 4PRESET5 on the EMC analyzer, then wait for the preset routine to nish. Set the EMC analyzer by pressing the following keys:
4FREQUENCY5 Band Lock
The HP 8594EM does not need to be band locked.
4FREQUENCY5 300 4MHz5
CF STEP AUTO MAN (MAN) 100 4MHz5
4SPAN5 0 4Hz5
4BW5 IF BW AUTO MAN (MAN) 30 4kHz5
3. On the EMC analyzer, press the following keys:
4MKR5
4AUX/USER5 Track Gen SRC PWR ON OFF (ON) 102 4dBV5
4.On the EMC analyzer, press TRACKING PEAK . Wait for the PEAKING message to disappear.
5.Zero and calibrate the measuring receiver and 100 kHz to 2.9 GHz power sensor in log mode (power reads out in dBm), as described in the measuring receiver operation manual. Enter the power sensor 300 MHz Cal Factor into the measuring receiver.